● Case No. : MEP-2230-003
● Device : Olympus Colonoscope CF-HQ290L
● Symptom : 1. Max Angulation OOS. 2. Abnormal mechanical noise during angulation. 3. Loss of angulation control in specific range.
● Repair Details : Relubrication / Internal realignment / Coil pipe Assy. replacement
● Result & QC : QC Passed