CF-Q260AL ‘MEP-2111-062’
2026-05-04
● Case No. : MEP-2111-062
● Device : Olympus Colonoscope CF-Q260AL
● Symptom :
1. Slack in the angulation mechanism
2. Overall deterioration and buckling of the insertion tube.
3. Leakage detected at Switch #1
4. Leakage detected at Biopsy Channel
5. Breakage of the angulation coil pipe assembly
● Repair Details :
– Insertion tube unit replacement
– Control body section relubrication and mechanical backlash adjustment
– Switch #1 rubber cover replacement
– Coil pipe Assy. replacement
● Result & QC : QC Passed